Method of Improving the Measurement Accuracy of Magnetic Material Characteristic for Open Circuit Test Specimen

Qiming Zhao,Hua Zhang,Yunqi Shangguan,Jun Zou,Jiansheng Yuan
DOI: https://doi.org/10.1109/icems.2017.8056181
2017-01-01
Abstract:The standard measurement method for soft magnetic materials can not measure the magnetic parameters in deep saturation. Using the measurement equipment for hard magnetic materials, which has strong excitation ability, the parameters in deep saturation can be measured, whereas the measurement error is very large. Based on the analysis of the error sources, this paper proposed a method to improve the measurement accuracy. First, the magnetic field probe near the test specimen is moved to the area where is away from the specimen. Then, the relationship array between the external measurement point and the internal magnetic field under different magnetic permeability is established by numerical simulation. Finally, the iterative optimization method determines the internal field values from the external measurements based on the relationship array. The method try to deduce B and H to the same point within the material as much as possible, thus forming a relatively accurate hysteresis loop of the material.
What problem does this paper attempt to address?