A high-accuracy measurement method of glucose concentration in interstitial fluid based on microdialysis

Dachao Li,Qingmei Xu,Yu Liu,Ridong Wang,Kexin Xu,Haixia Yu
DOI: https://doi.org/10.1088/1361-6501/aa87e4
IF: 2.398
2017-01-01
Measurement Science and Technology
Abstract:A high-accuracy microdialysis method that can provide the reference values of glucose concentration in interstitial fluid for the accurate evaluation of non-invasive and minimally invasive continuous glucose monitoring is reported in this study. The parameters of the microdialysis process were firstly optimized by testing and analyzing three main factors that impact microdialysis recovery, including the perfusion rate, temperature, and glucose concentration in the area surrounding the microdialysis probe. The precision of the optimized microdialysis method was then determined in a simulation system that was designed and established in this study to simulate variations in continuous glucose concentration in the human body. Finally, the microdialysis method was tested for in vivo interstitial glucose concentration measurement.