Iris-Claw Intraocular Lens And Scleral-Fixated Posterior Chamber Intraocular Lens Implantations In Correcting Aphakia: A Meta-Analysis

Wu Jing,Liang Guanlu,Zheng Qianyin,Li Shuyi,He Fengying,Liu Jian,Xu Wen
DOI: https://doi.org/10.1167/iovs.16-21226
2017-01-01
Abstract:PURPOSE. A meta-analysis to compare iris-claw intraocular lens (IC-IOL) and scleral-fixated posterior chamber intraocular lens (SF-PCIOL) implantations in correcting aphakia without sufficient capsular support.METHODS. Eligible studies were collected through PubMed, Web of Science, Embase, and the Cochrane library. The pooled relative risks (RR), pooled standardized mean difference (SMD), and their 95% confidence interval of the eligible studies were then calculated. Seven studies met our inclusion criteria, involving 232 and 158 eyes in IC-IOL and SF-PCIOL groups, respectively.RESULTS. The pooled SMD of the mean postoperative corrected distance visual acuity (CDVA) (logMAR) was -0.25. The pooled RR of the eyes achieving 20/40 or better postoperatively was 1.16. The pooled SMD of the surgical time was -2.97. The pooled RR of the surgical complications was 0.86. The pooled RR of IOL dislocation, retinal detachment (RD), and cystoid macular edema (CME) between the two groups were 0.22, 0.63, and 0.64.CONCLUSIONS. Implantation of IC-IOL has a more simple procedure and shorter learning curve than SF-PCIOL implantation in correcting aphakia without sufficient capsular support.
What problem does this paper attempt to address?