Chemiluminescence Immunoassay for Cardiac Troponin T by Using Silver Nanoparticles Functionalized with Hemin/g-Quadruplex DNAzyme on a Glass Chip Array

Chen Zong,Duoduo Zhang,Hua Yang,Shumei Wang,Ming Chu,Ping Li
DOI: https://doi.org/10.1007/s00604-017-2331-z
IF: 5.7
2017-01-01
Microchimica Acta
Abstract:The authors describe an array for chemiluminescence (CL) based determination of cardiac troponin T (cTnT), an important cardiovascular disease marker. The tracing tag consists of silver nanoparticles (AgNPs) loaded with guanine-rich DNA sequences and detection antibody in a high numerical ratio. The loaded AgNPs were then reacted with hemin to form a hemin/G-quadruplex DNAzyme. A disposable immunosensor array was fabricated by immobilizing capture antibody on corresponding sensing sites on a glass chip. Once a sandwich immunocomplex is formed on the array, the tracing tag catalyzes the CL reaction of the luminol-p-iodophenol and H2O2 system to produce a CL signal, which is collected by a CCD camera. An intuitive CL image is obtained containing all of the spots on the array. Under optimal conditions, the method shows a wide linear range over 4 orders of magnitude (from 0.003 to 270 ng·L−1), a detection limit down to 84 fg·L−1, and a throughput as high as 44 tests·h−1. The results obtained with serum samples are in acceptable agreement with reference values. The AgNP-based tracing tag as well as the immunoassay method shows a promising potential for point-of-care testing for the early clinical diagnosis of cardiovascular disease.
What problem does this paper attempt to address?