A Wireless Wearable Surface Functional Electrical Stimulator

Hai-Peng Wang,Ai-Wen Guo,Yu-Xuan Zhou,Yang Xia,Jia Huang,Chong-Yao Xu,Zong-Hao Huang,Xiao-ying Lu,Zhi-Gong Wang
DOI: https://doi.org/10.1080/00207217.2017.1312708
2017-01-01
International Journal of Electronics
Abstract:In this paper, a wireless wearable functional electrical stimulator controlled by Android phone with real-time-varying stimulation parameters for multichannel surface functional electrical stimulation application has been developed. It can help post-stroke patients using more conveniently. This study focuses on the prototype design, including the specific wristband concept, circuits and stimulation pulse-generation algorithm. A novel stimulator circuit with a driving stage using a complementary current source technique is proposed to achieve a high-voltage compliance, a large output impedance and an accurate linear voltage-to-current conversion. The size of the prototype has been significantly decreased to 17x7.5x1cm(3). The performance of the prototype has been tested with a loaded resistor and wrist extension/flexion movement of three hemiplegic patients. According to the experiments, the stimulator can generate four-channel charge-balanced biphasic stimulation with a voltage amplitude up to 60V, and the pulse frequency and width can be adjusted in real time with a range of 100-600s and 20-80Hz, respectively.
What problem does this paper attempt to address?