An Individually Designed Custom-Made Mandibular Condyle Prosthesis Using Selective Laser Melting: Finite Element Analysis

X. Xu,D. Luo,C. Guo,Q. Rong
DOI: https://doi.org/10.1016/j.ijom.2017.02.806
IF: 2.986
2017-01-01
International Journal of Oral and Maxillofacial Surgery
Abstract:Background: In adults, alloplastic reconstructions are a predictable option for reconstruction of mandibular condyle defects. Additive manufacturing techniques have recently been developed to fabricate structures for tissue engineering, which could be used for individually designed condyle prosthesis. Objectives: A novel individually designed selective laser melting (SLM) three-dimensional printing custom-made mandibular condyle prosthesis was proposed. Methods: The head of an adult beagle dog was scanned to get the computed tomography image used as the condyle reconstruction model. The mandible, bilateral discs and bilateral temporal bone were constructed using ANSYS software. The material of the mandibular condyle prosthesis was titanium-6 aluminum-4 vanadium (Ti-6Al-4V). An optimised quadrihedron open-porous scaffold with combined connection structures, an inlay rod and an onlay plate, between the prosthesis and remaining mandible was designed as the mandibular condyle prosthesis. The biomechanical behaviours of the mandible and contralateral temporomandibular joint (TMJ) in two models, intact mandible and mandible with prosthesis, were compared after the bite force had been applied. Findings and Conclusion: The magnitude of the compressive strain on the condyle neck with the prosthesis was low, except around the screws, while the magnitude of the strain on the scaffold-bone interface was relatively high. Reconstruction of the mandibular condyle with prosthesis would affect the contralateral TMJ. The finite element analysis model of the mandibular condyle prosthesis also showed the prosthesis could transfer the bite force appropriately.
What problem does this paper attempt to address?