Evaluation of Constrained-TV-Minimization Image Reconstruction from Reduced PET Data

Zheng Zhang,Jinghan Ye,Buxin Chen,Amy E. Perkins,Andreia Trindade,Pedro Rodrigues,Chien-Min Kao,Emil Y. Sidky,Xiaochuan Pan
2015-01-01
Abstract:1830 Objectives The aim of this study to evaluate the quality of phantom images reconstructed by use of constrained-total-variation (TV)-minimization method from reduced PET acquisition simulating 50% of scintillation crystals. Methods We employ a digital PET scanner to collect data using an IEC phantom, which contains 6 spheres of 10, 13, 17, 22, 28, and 37 mm diameter. The activity in the smallest four spheres is 4 times the background and zero in the other two. The total activity at the beginning of data acquisition is 53 MBq. We investigate a sparse PET configuration, which is to remove odd-numbered tiles in the odd-numbered rings, and even-numbered tiles in the even-numbered rings. We use the adaptive-steepest-descent (ASD)-projection-onto-convex-set (POCS) algorithm to solve a constrained-TV-minimization program, which is to seek an image with minimum image TV within a feasible solution set. We reconstruct images from full data by using the row-action maximum likelihood algorithm (RAMLA) as references. Images are evaluated by use of signal to noise ratio (SNR) of three larger hot spheres, and the selected ROIs are 10 mm in diameter. Results The SNR measurements are shown in the table. For the hot spheres of size 22 and 17 mm, the SNR in the ASD-POCS reconstructions are higher than or comparable to that in the RAMLA result. For the hot sphere of size 13 mm, the SNR in the ASD-POCS reconstructions are slightly lower than that in the RAMLA image. It is worth noting that, the ASD-POCS reconstruction from sparse data, although degraded in quality as a result of fewer events, can still reveal the smallest hot sphere of size 10 mm. Conclusions Image quality of the ASD-POCS reconstructions from full data is comparable to, or slightly better than RAMLA reconstructions in terms of SNR of hot spheres. ASD-POCS reconstructions from data with 50% crystal reduction possess quality which may potentially yield clinical utilities. Research Support This work was supported in part by NIH R01 Grant Nos. CA120540, CA158446, and EB000225, and a Philips research grant.
What problem does this paper attempt to address?