Clinical efficacy of a locator for puncture under CT examination

Bing HUANG,Jian-guo GUO,Min YAN,Ming YAO,Hao-qiang CAO,Xu-yan ZHOU,Jian HOU,Ya-ping LU,Jian-liang SUN
DOI: https://doi.org/10.3760/cma.j.issn.0254-1416.2012.09.018
2012-01-01
Abstract:We designed a locator for puncture under CT examination.Thirty-three patients of both sexesaged 16-76 yr weighing 46-80 kg undergoing bilateral thoracic or lumber sympathetic nerve block under CT examination were enrolled in this study.One side was punctured under the guidance of the locator,while the other side was punctured by conventional technique.The time consumed during puncture,CT scan adjusting time,the deviation of the needle from the targets and the incidence of puncture complications were recorded.The results showed that compared with the conventional technique,with the locator the time consumed during puncture was significantly shorter,the number of attempts and the deviation of the needle from the target were significantly reduced.The procedure was successfully performed on both side in all patients without serious complications such as hemothorax and pneumothorax.
What problem does this paper attempt to address?