High Spatiotemporal Resolution Magnetic Resonance Thermometry Based on Partially Separable Theory

Caiyun Shi,Guoxi Xie,Xiaoyong Zhang,Shi Su,Yongqin Zhang,Yanchun Zhu,Chao Zou,Xin Liu
DOI: https://doi.org/10.1166/jmihi.2017.2004
2017-01-01
Journal of Medical Imaging and Health Informatics
Abstract:It is very difficult to achieve high spatiotemporal resolution MR thermometry (MRT) due to the slow sampling speed of magnetic resonance (MR). To address this issue, we presented a novel method to improve the spatial and temporal resolutions of MRT by exploiting data redundancy based on partially separable (PS) theory. After high spatioternporal resolution images with accurate phase information were reconstructed by the PS theory, high spatiotemporal resolution MRT achieved using a standard proton resonance frequency (PRF) shift technique. The simulation, phantom and in vivo experiment results showed that the proposed method could accurately estimate dynamic temperature change with high spatiotemporal resolution. In the in vivo experiment, the temporal resolution can reach to 250 ms in the case of spatial resolution with 1.2 x 1.2 mm(2). These results demonstrated the feasibility of the proposed method for high spatiotemporal resolution MRT.
What problem does this paper attempt to address?