An Electrode Design for Droplet Dispensing with Accurate Volume in Electro-Wetting-based Microfluidics

Wei Wang,Jianfeng Chen,Jia Zhou
DOI: https://doi.org/10.1063/1.4954195
IF: 4
2016-01-01
Applied Physics Letters
Abstract:Electro-wetting-on-dielectric actuation enables droplets, the basic units of digital microfluidics, to be manipulated on a two-dimensional surface, providing a versatile platform for chemical assays or multi-step operations at micro- or nano-scale. In this paper, we carry out characterization experiments to demonstrate an electrode design to improve the reproducibility of on-chip droplet generation with no extra external apparatus. The overall reproducibility for a sequence of droplets dispensed consecutively from a non-refilling reservoir can be limited within ±0.5%. Results from the repetition of 1000 iterations offer the long-term reproducibility in the range of ±1%, indicating its validity in practical applications.
What problem does this paper attempt to address?