Ultra-Wide Detectable Concentration Range of Gmr Biosensors Using Fe3o4 Microspheres

Jie Xu,Qiang Li,Weihua Zong,Yongcheng Zhang,Shandong Li
DOI: https://doi.org/10.1016/j.jmmm.2016.05.059
IF: 3.097
2016-01-01
Journal of Magnetism and Magnetic Materials
Abstract:Exchange-biased GMR sensors were employed for biodetection using a DC in-plane measuring method and a magnetic label of Fe3O4 microspheres. It was revealed that an ultra-wide concentration span covering five orders from 10ng/mL to 1000μg/mL was achieved in a home-made biodetection device. The concentration x dependence of output voltage difference |ΔV| between with and without magnetic labels, exhibits nonlinear futures, which undergoes two functions depending on the concentration region. For the low concentration region from 10ng/mL to 10μg/mL, a logarithmic relation of |ΔV|=26.3lgx+91.4 fits well, while for the high concentration region, a negative exponential function of |ΔV|=3113(1−e−x/250) describes the |ΔV|~x relation better. For the former, the “coffee ring” effect, formed during the solvent evaporation, was considered as the main reason for the nonlinear relation. While for the latter with high concentration, the overlap among the particles and the enhanced interaction of the magnetic dipole were responsible for the nonlinear |ΔV|~x relationship. Moreover, the calculated detectable concentration limit is agreed well with the experimental data.
What problem does this paper attempt to address?