Oxygen concentration detection system based on TDLAS and online elimination of background noise

Tao PANG,Hua XIA,Bian WU,Zhi-rong ZHANG,Yu WANG,Xiao-juan CUI,Feng-zhong DONG
DOI: https://doi.org/10.16136/j.joel.2015.03.0479
2015-01-01
Abstract:Oxygen concentration detection plays an important role in industrial process control.In this paper,tunable diode laser absorption spectroscopy(TDLAS)is used for online monitoring of oxygen concentration.The experimental system employs vertical cavity surface emitting lasers(VCSEL)with output wavelength of 760 nm as light source and adopts the radio-type structure with optical path of 60 cm.Continuous measurement for 72 hours shows that the system has a high accuracy in short time,but slow drift comes out when the time is longer than 4hours.Based on the analysis of first harmonic signal and second harmonic signal,this paper argues that the changes of background noise causes measurement error,and the composition and characteristics of background noises of the system are carried out and discussed too.After analyzing the system signal from the perspective of frequency domain,we propose that under the premise of the amplitude-frequency characteristic of the system′s background noise keeping stable,cross-correlation algorithm could be applied to deduce the background noise in real-time.The test of 5% oxygen for 60 hours points out that after data processing,the maximum relative error is reduced from 3.2%to 1.82%.This indicates that the proposed method could effectively improve the system′s performance.
What problem does this paper attempt to address?