Electrogenerated Chemiluminescence Detection Chlortetracycline Based on Energy Transfer

刘丽君,黄赣生,刘利民
2003-01-01
Abstract:In this paper, it was found that the weak ECL signal, generated from electrooxidation fluorescein on the surface of the carbon- printed electrode, was greatly enhanced by the chlortetracycline. Based on this observation, a selective ECL method for the chlortetracycline was developed in this paper. Under the optimum experimental conditions developed, the detection limit of chlortetracycline was 610-7g/ml, and the linear range for deletion chlortetracycline was 2.0×10-6-5.0×10-5g/ml with the relative standard deviation of 3.1%[c=1×10-5g/ml,n=11].
What problem does this paper attempt to address?