Extraction method of microfluidic PCR fluorescent signal for gene detection

赵树弥,朱灵,李阳,朱灿灿,李志刚,张龙,邓国庆,王安,刘勇
DOI: https://doi.org/10.16136/j.joel.2014.06.031
2014-01-01
Abstract:To resolve the problems of polymerase chain reaction(PCR)chambers identification and fluorescence signal extraction in microfluidic real-time PCR system,a data extraction method based on algorithm integration is proposed.First,we extract the target region which contains many PCR chambers from the CCD image.Background and the PCR chambers are separated by the minimum error threshold method.The grid method is used to extract a single reaction chamber area.The edge of each reaction chamber is identified with the wavelet transform and the improved Canny operator.Second,the average gray value of each PCR chamber is calculated and is used as the expression of the gene amplification.A chip with four chambers loading with PUC-18gene is tested by our microfluidic real-time system.The results show that our method can identify PCR chambers effectively and calculate the intensity of the fluorescence signal rapidly.The repeatability of PCR amplification curves is improved by eliminating the influence of the bright and dark spots.The accuracy of gene amplification analysis is also enhanced.
What problem does this paper attempt to address?