Charateristic of dose distribution and its clinical application in stereotactic radiotherapy practice

郎锦义,李建,吴大可,张德康,李黎
DOI: https://doi.org/10.3969/j.issn.1005-202X.2003.01.003
2003-01-01
Abstract:Purpose: To measure clinical dosimetric parameters and to evaluate the accuracy of isocenter dose distribution in circular collimater fields in stereotactic radiotherapy for correctly using of stereotactic radiotherapy in clinical practice. Materials and Methods: Polystyrene phantom, Anthropomorphic head phantom and film-checking technique were used to measure the parameters of off-axis ratio, penumbra, steepness and flatness of isodose curves under static, single-target, and 2-targets-isocenters moving fields condition in BrainLAB X-knife system. Conclusion: Our parameters obtained from above-mentioned different measurement methods demonstrated that to understand thoroughly the variations of clinical dosimetric parameters at the isocenter of circular collimeter fields is one of the key factors for performing correctly the stereotactic radiotherapy techniques in clinical practice, such as, when you choose multi-targets X-knife therapy method for the nasopharyngael carcinoma. Some suggestions were offered according to the results of our test measurements for guiding the correct application of stereotactic radiotherapy in clinical practice.