Using the 64-Slice Perfusion CT to Evaluate the Oxygen Tension(pO2) in the Rabbit VX2 Tumor Model: an Experimental Study

孙昌进,王光辉,李超,李涛,罗云秀,吕海波,张德康,李彦,黄建鸣,肖明勇,阴俊,于金明,郞锦义
DOI: https://doi.org/10.3969/j.issn.1674-0904.2013.01.003
2013-01-01
Abstract:Objective:To investigate the role of the 64-slice perfusion CT in the evaluation of the oxygen tension(pO_2 ) in the rabbit VX_2 tumor model.Methods:Forty-five rabbit VX_2 brain tumor model established successfully were examined with 64-detector row CT.Tumor specimens were assessed for the oxygen tension(pO_2 ),perfusion,blood volume(BV),peak enhancement intensity(PEI) and time to peak(TTP),and Pearson correlation coefficients were conducted to represent the relationships between the perfusion parameters and pO_2 of the tumor.pO_2 was measured by oxygen-sensitive electrodes guided by perfusion CT images.Results:Mean values for perfusion,BV,PEI,TTP and pO_2 of the 45 tumors were 27.102±26.723ml/min,22.1 96±13.680ml/100g,43.456±28.73 HU, 38.823±14.759 sec,and 15.981±14.815mmHg,respectively.BV.PEI,TTP were not significant correlated with pO_2(r =0.271、0.253、- 0.18,P 0.05),whereas positively correlation was found between perfusion with pO_2(r = 0.673,P = 0.00).Conclusion: The perfusion value from 64-slice spiral CT perfusion imaging might to have ability to evaluate the tumor pO_2