Multichannel electrochemical immunosensor arrays for the detection of human serum total prostate-specific antigen

Pei-jun MA,Zhen-hua LI,Yan-zhi DOU,Wang-ping DENG,Jian-xin LV,Shi-ping SONG,Chun-hai FAN
DOI: https://doi.org/10.13602/j.cnki.jcls.2014.08.01
2014-01-01
Abstract:Objective To establish a simple,sensitive and specific method to detect human serum total prostate-specific antigen( tPSA) based on multichannel electrochemical immunosensor arrays( ECISA) and evaluate its performance. Methods Anti-PSA antibody was immobilized on a disposable 16-channel screen-printed carbon electrode array by an antibody vacuum assembly strategy.Then,the established multichannel ECISA combined with sandwich immununoassay,was used to detect the levels of tPSA in 185 human serum samples and the obtained results were compared with those detected by electrochemiluminescence immunoassay( ECLIA).Results When the established multichannel ECISA was used to detect serum tPSA level,its linear range was 0. 05-50 ng /mL for the tPSA standard curve,the limit of detection was 10 pg /mL,the limit of quantification was 61 pg /mL,and the intra-assay coefficient of variation( CV) was 8. 00%-8. 50%. There was good correlation between the results of ECISA and those of ECLIA( r2= 0. 941,P 0. 01). Conclusion The established multichannel ECISA has high sensitivity and specificity,and good correlation with ECLIA for the detection of serum tPSA,which may be a potential tool for the detection of tPSA and other tumor markers.