Establishment of the Standard Curve of Detecting EFG by Indirect Competitive Enzyme-Linked Immunosorbent Assay (Cielisa)

李雪玲,吴学诗,赖锦昌,李冰,张美英,夏勇,王一飞
DOI: https://doi.org/10.3969/j.issn.1000-9965.2003.06.007
2003-01-01
Abstract:Aim: To establish a quick quantitative ELISA for detecting EGF. Methods: The rhEGF was the coating antigen and competitive antigen. They react to the limited amount of polyclonal antibody against rhEGF. Results: The optimal concentration of the coating antigen was 1 μg/mL, The dilutions of polyclonal antibody against rhEGF and sheep anti-rabbit IgG were 1/(10 000) and 1/(3 000) respectively. Quantization of the rhEGF was linear from 0.5~16 ng/mL, with the detection limit being 0.5 ng/mL. The coefficients of variation of intra-assay and inter-assay were 6.21% and 7.42% respectively. Then, the regression equation was y=-13.65 ln(x)+81.554 (R~2=0.997) and there was a standard curve. Conclusion: The indirect quick competitive enzyme-linked immunosorbent assay was established.