Clinical Evaluation of Circumferential Ultrasound Ablation of Pulmonary Veins for Patients with Paroxysmal Atrial Fibrillation

马长生,刘兴鹏,刘旭,王建安,董建增,孙英贤,卢才义,陈慧敏,郭成军,方冬平
DOI: https://doi.org/10.3760/j:issn:0253-3758.2003.07.003
2003-01-01
Abstract:Objective To evaluate the efficacy and safety of ultrasound balloon ablation of pulmonary veins for patients with paroxysmal atrial fibrillation (PAF).Methods Circumferential ultrasound ablation of both upper pulmonary veins (PVs) and left inferior PV was performed in 47 patients with frequent symptomatic and drug refractory PAF.The procedural endpoint was PVs isolation.Results A total of 137 PVs were ablated, and 95 of them achieved the procedural endpoint.After a mean follow-up of 3-19 (11.7±5.1) months in 44 patients, PAF was completely eliminated in 18 patients (40.9%) without antiarrhythmic drug .Complications included right phrenic nerve paralysis in 2 patients,transient ST segmental elevation in inferior leads in 4 and severe vagus reflex in 2.No evidence of narrowing of PVs was seen with repeat venography after ablation.Conclusions Ultrasound balloon ablation system can isolate multiple PVs with an acceptable results.Developing special design of ultrasound balloon catheter and ultrasound energy delivering system at the PVs ostium with a high temperature ( 60-65 ℃) may be important to maximize success rate and reduce the complications.