Design and biomechanical study of cervical vertebra titanium plate-interbody fusion cage

柏传毅,王坤正,魏美钢,党晓谦,王春生,时志斌
DOI: https://doi.org/10.3969/j.issn.1004-406X.2003.10.009
2003-01-01
Abstract:Objective:To enhance anterior cervical verteb ra internal fixation device,a new cervical vertebra in-ternal fixation device n amed titanium plate-interbody fusion cage was devised.After the biomechanical st udy was performed,the possibility of clinical application was evaluated.Method:S ome functional spinal unit was fixed respectively with titanium plate,fusion cag e and new device designed by us.The mechanical experimental study including flex ion,extension,torsion and lateral bending were perfomed and a comparative study was made.Result:As to the mechanical performance,fusion cage was unsatisfied und er extension test and titanium plate was in the same condition under distortion test.However,new device was satisfied in every test items.Conclu-sion:Both simp le titanium plate fixaiton and simple fusion cage fixation have a biomechanical weak point,but some complementation existed between them,the titanium plate-inte rbody fusion cage has occupied their ad-vantages.The superiority could be obser ved obviously in biomechanical test.The new device can meet the clin-ical appli cation of cervical disorders.
What problem does this paper attempt to address?