Automated Enzymatic Determination for Serum Total and Free L-carnitine and Its Clinical Application

陈肖燕,谢鑫友,费春荣
DOI: https://doi.org/10.3760/j:issn:1009-9158.2003.12.019
2003-01-01
Abstract:Objective To establish an automated spectrophotometic assay for serum total and free L-carnitine with enzymatic method(CAT-BTND). Methods To optimize the reaction conditions of hydrolysis,pH and L-carnitine Acetyl-transferase (CAT)concentration in the assay,to evalute the assay method and to measure the serum L-carnitine of health adults ( n =63) and children ( n =56). Results The calibration curve for L-carnitine was linear between 4 81 and 150 μmol/L( r =0 999).The inter and intra average CV were 5 52% and 5 90% in free carnitine,5 78% and 6 14% in octanoylcarnitine.The mean recovery rate was 98 5% and 98 4% respectively.This method was free from interference by bilirubin(≤120 μmol/L),hemoglobin(≤7 g/L),and ascorbate(≤500 μmol/L).Total and free carnitine and acylcarnitine reference range: adults (48 5±17 7) μmol/L , (39 2±14 2) μmol/L and (9 3±5 3) μmol/L(male, n =32),(42 7±17 3) μmol/L, (32 4±18 0) μmol/L and (10 5±5 5) μmol/L respectively(female, n =31).Children (age between 9~12 years old) (57 3±19 3) μmol/L , (45 1±14 4) μmol/L and (12 3±10 4) μmol/L(male, n =31), (59 4±15 7) μmol/L, (46 6±12 6) μmol/L and (12 8±7 8) μmol/L respectively(female, n =25). Conclusion The procedure is rapid,accurate and automatable,and could be used to diagnose L-carnitine deficiency,and to estimate and monitor the effect of treatments.
What problem does this paper attempt to address?