Dislocation of Electrode During or after Deep Brain Stimulation

占世坤,孙伯民,沈建康
DOI: https://doi.org/10.3969/j.issn.1009-122x.2004.10.003
2004-01-01
Abstract:Objective To evaluate the phenomenon of dislocations of the placed electrode in patients with advanced Parkinson disease undergoing deep brain stimulation. Methods The electrode positions were compared with the preoperative planned targets by MRI or X ray. Results Dislocation of electrode was found in 5 cases, it happened during operation in 2 cases, 3 cases had good external stimulation effects and the dislocation of the placed electrode compared with its planed target point was ruled out by intraoperative image, but MRI showed dislocation of electrode in 4 weeks after the operation, 2 cases of them showed 4mm deeper than preoperative planned targets, and 1 case showed 6mm deeper. Conclusions The dislocation of electrode was an important factor that made the chronic stimulation effect worse in some patients who had had good external intraoperative stimulation effects, and should be actively prevented. Postoperative MR image could verify the dislocation of electrode and help processing.