Measurement Of Photoelectron Yield Of The Cdex-10 Liquid Argon Detector Prototype

Qing-Hao Chen,Qian Yue,Jian-Ping Cheng,Ke-Jun Kang,Yuan-Jing Li,Shin-Ted Lin,Chang-Jian Tang,Hao-Yang Xing,Xun-Zhen Yu,Ming Zeng,Jing-Jun Zhu
DOI: https://doi.org/10.1088/1674-1137/40/11/116002
IF: 2.944
2016-01-01
Chinese Physics C
Abstract:The China Dark Matter Experiment (CDEX) is a low background experiment at China Jinping Underground Laboratory (CJPL) designed to directly detect dark matter with a high-purity germanium (HPGe) detector. In the second phase, CDEX-10, which has a 10 kg germanium array detector system, a liquid argon (LAr) anti Compton active shielding and cooling system is proposed. To study the properties of the LAr detector, a prototype with an active volume of 7 liters of liquid argon was built and operated. The photoelectron yields, as a critically important parameter for the prototype detector, have been measured to be 0.051-0.079 p.e./keV for 662 keV gamma rays at different positions. The good agreement between the experimental and simulation results has provided a reasonable understanding and determination of the important parameters such as the surviving fraction of the Ar-2(*) excimers, the absorption length for 128 nm photons in liquid argon, the reflectivity of Teflon and so on.
What problem does this paper attempt to address?