Accuracy of Fetal Lung Volume Assessed by Three-dimensional Sonography

zhang bo,yang taizhu
DOI: https://doi.org/10.3969/j.issn.1002-0101.2011.02.025
2011-01-01
Abstract:Objective To assess the value of 3D ultrasound in fetal lung volume measurement using VOCAL technique,and to determine its accuracy and reproducibility.Methods Fetal lung volumes were measured using VOCAL technique in 14 cases with pulmonary hypoplasia and in 17 controls 1-3 days before termination.Postmortem autopsy was conducted within 6 hours after labor and the lung volume was achieved by water displacement.Results The correlation between measured volume and postmortem examination was significant,with correlation coefficient 0.80 in pathological cases and 0.96 in controls.The mean relative error,bias,95%limits of agreement were,respectively, -2.31%,-0.19 cm~3 and(-3.59-3.20) cm~3 in cases with pulmonary hypoplasia and -1.06%,-0.37 cm~3 and (-6.12-5.38)cm~3 in controls.Good intraobserver and interobserver reproducibility were achieved,both in pathological cases and in controls.Conclusions Prenatal 3D ultrasound can estimate accurately fetal lung volume using VOCAL technique,both in normal and pathological situation,which conduces to evaluation of its physiological development.