Experimental evaluation of the gel antiglobulin test

chen xuejun,xu xingqiang,jin xiaobo,cheng yongzhang
DOI: https://doi.org/10.3969/j.issn.1673-8640.2004.04.027
2004-01-01
Abstract:Objective To evaluate the characteristics of the gel antiglobulin test (GAT). Methods A series of tests were used to analyse some characteristics of GAT(LISS/Coombs card, DiaMed), with commercialized anti D(IgG) as simulation samples. Results Elimination of washing step did not affect result of GAT. The strength of positive reaction (1∶64 anti D and 0.8% RBCs) began to decrease slowly when 40 μl AB plasma was added into LISS/Coombs card. Compared with ID diluent2, other diluents (NS, LIM) were not suitable for GAT. The Highest titer of anti D(IgG) measured by GAT was 1∶256, and the sensitivity of GAT was higher than that of traditional antiglobulin test, and approximately equal to manual polybrene test. Conclusions GAT has some specialities such as no washing steps, efficient ID diluent2, high sensitivity, and so on. It is very important to follow DiaMed’s operation manual, otherwise false results would be obtained.