An Automated Dose Verification Software for Brachytherapy

Xianliang Wang,Pei Wang,Churong Li,Zhangwen Wu,Chengjun Gou,Jie Li,Shengwei Kang,Qing Hou
DOI: https://doi.org/10.5114/jcb.2018.79396
2018-01-01
Journal of Contemporary Brachytherapy
Abstract:Purpose: To report an implementation method and the results of independent brachytherapy dose verification software (DVS). Material and methods: The DVS was developed based on Visual C++ and adopted a modular structure design. The DICOM RT files exported from a treatment planning system (TPS) were automatically loaded into the DVS. The DVS used the TG-43 formalism for dose calculation. A total of 15 cervical cancer patients who underwent brachytherapy were retrospectively selected to test the DVS. Dosimetric parameters and gamma analysis (0.1 cm, 5%) were used to evaluate the dose differences between the DVS and the TPS. Results: Compared with the TPS dose, the gamma pass rates of the dose calculated by the DVS were higher than 98%. For the clinical target volume (CTV), the dosimetric differences were less than 0.63% for D-90(%) and D-100%. For the bladder, rectum, and sigmoid, the agreement of D-0.1cc, D-1cc, and D-2cc were within a 0.78% level. Conclusions: With minimal human-computer interactions, the DVS can verify the accuracy of doses calculated by the TPS.