An Optimized LC-MS/MS Method for Determination of HYNIC-3PRGD2, a New Promising Imaging Agent for Tumor Targeting, in Rat Plasma and Its Application.

Liyun Niu,Jian Li,Ruolan Gu,Zhuona Wu,Xiaoxia Zhu,Hui Gan,Baiping Ma,Bing Jia,Fan Wang,Zhiyun Meng,Xinyu Wang,Guifang Dou
DOI: https://doi.org/10.1016/j.jchromb.2018.07.026
IF: 3.318
2018-01-01
Journal of Chromatography B
Abstract:HYNIC-3PRGD(2) is used to prepare a new Tc-99m-radiolabeled tracer. HYNIC-3PRGD(2), which has a high binding affinity for the integrin alpha(v)beta(3) due to its special structure, has become a promising tumor imaging agent for diagnosis and monitor of the clinical response to therapeutic effects of anti-tumor agents. Here, we developed and validated a method for determination of HYNIC-3PRGD(2) concentration in rat plasma using ultra-high performance liquid chromatography-tandem mass spectrometry system. Following sample extraction by methanol precipitation, satisfactory separation through chromatography was achieved on an hydrophilic reverse-phase C-18 column AQ (2.1 mm x 100 mm, 2.7 mu m) at a flow rate of 0.2 mL.min(-1) with an gradient elution using mobile phase consisting of ultrapure water and acetonitrile fortified with 0.1% formic acid respectively. The calibration curve was developed over a linear range of 3.125-100 ng.min(-1) with the lower limit of quantification of 3.125 ng.min(-1). The HYNIC-3PRGD(2) and its internal standard c(RGDfK)(RK5) were detected and quantified with the multiple reaction monitoring (MRM) mode on a triple-quadrupole tandem mass spectrometer. This method was successfully validated and applied for pharmacokinetic evaluation of HYNIC-3PRGD(2) during preclinical experiments.
What problem does this paper attempt to address?