Collecting System Percutaneous Access Using MR-assisted Guidance:First Human Phantom Experience

H Zhu,N C Li,Z C Zhao,H G Wang,B Zhang,Y Q Na
DOI: https://doi.org/10.3760/cma.j.issn.0376-2491.2018.24.014
2018-01-01
Abstract:Objective Precise renal puncture is an essential but challenging step for successful percutaneous nephrolithotomy.We evaluated the efficiency of a novel real-time navigation system using Mixed Reality(MR)technology for human phantom kidney puncture.Methods One human kidney phantom underwent MR-assisted percutaneous collecting system puncture.Two punctures were performed by each of 6 surgeons in the randomize selected upper,middle or lower calyces.Outcome measurements were the number of attempts for renal puncture,the time needed to evaluate the trajectory and to perform percutaneous puncture.Results A total of 12 punctures were performed successfully using MR-Assisted Guidance.Median evaluation time and renal puncture time for the selected calyces was 13(range 11 to 19)and 19 seconds(range 15 to 44),respectively.One or Two attempts were needed to achieve a successful renal puncture for all of the surgeons.Conclusions The proposed MR-assisted guidance solution for renal collecting system puncture proved to be accurate,simple and quick.The inherent limitations of traditional X-ray and ultrasonic technology can be overcome.
What problem does this paper attempt to address?