Quantitative Characterization of Turbidity by Radiative Transfer Based Reflectance Imaging.

Peng Tian,Cheng Chen,Jiahong Jin,Heng Hong,Jun Q. Lu,Xin-Hua Hu
DOI: https://doi.org/10.1364/boe.9.002081
IF: 3.4
2018-01-01
Biomedical Optics Express
Abstract:A new and noncontact approach of multispectral reflectance imaging has been developed to inversely determine the absorption coefficient of μ a , the scattering coefficient of μs and the anisotropy factor g of a turbid target from one measured reflectance image. The incident beam was profiled with a diffuse reflectance standard for deriving both measured and calculated reflectance images. A GPU implemented Monte Carlo code was developed to determine the parameters with a conjugate gradient descent algorithm and the existence of unique solutions was shown. We noninvasively determined embedded region thickness in heterogeneous targets and estimated in vivo optical parameters of nevi from 4 patients between 500 and 950nm for melanoma diagnosis to demonstrate the potentials of quantitative reflectance imaging.
What problem does this paper attempt to address?