Respiratory Motion-Resolved, Self-Gated 4D-MRI Using Rotating Cartesian K-space (ROCK): Initial Clinical Experience on an MRI-guided Radiotherapy System

F. Han,Z. Zhou,D. Du,Y. Gao,M. Cao,N. Shaverdian,J. Hegde,M. Steinberg,P. Lee,A. Raldow,D. Low,K. Sheng,Y. Yang,P. Hu
DOI: https://doi.org/10.1016/j.radonc.2018.04.029
IF: 6.901
2018-01-01
Radiotherapy and Oncology
Abstract:Purpose: To optimize and evaluate the respiratory motion-resolved, self-gated 4D-MRI using Rotating Cartesian K-space (ROCK-4D-MRI) method in a 0.35 T MRI-guided radiotherapy (MRgRT) system. Methods and materials: The study included seven patients with abdominal tumors treated on the MRgRT system. ROCK-4D-MRI and 2D-CINE, was performed immediately after one of the treatment fractions. Motion quantification based on 4D-MRI was compared with those based on 2D-CINE. The image quality of 4D-MRI was evaluated against 4D-CT. The gross tumor volumes (GTV) were defined based on individual respiratory phases of both 4D-MRI and 4D-CT and compared for their variability over the respiratory cycle. Result: The motion measurements based on 4D-MRI matched well with 2D-CINE, with differences of 1. 04 +/- 0.52 mm in the superior-inferior and 0.54 +/- 0.21 mm in the anterior-posterior directions. The image quality scores of 4D-MRI were significantly higher than 4D-CT, with better tumor contrast (3.29 +/- 0.76 vs. 1.86 +/- 0.90) and less motion artifacts (3.57 +/- 0.53 vs. 2.29 +/- 0.95). The GTVs were more consistent in 4D-MRI than in 4D-CT, with significantly smaller GTV variability (9.31 +/- 4.58% vs. 34.27 +/- 23.33%). Conclusion: Our study demonstrated the clinical feasibility of using the ROCK-4D-MRI to acquire high quality, respiratory motion-resolved 4D-MRI in a low-field MRgRT system. The 4D-MRI image could provide accurate dynamic information for radiotherapy treatment planning. (C) 2018 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?