Electrochemiluminescence Based Competitive Immunoassay for Sudan I by Using Gold-Functionalized Graphitic Carbon Nitride and Au/Cu Alloy Nanoflowers

Wanlu Chen,Xun Yao,Xinchun Zhou,Kang Zhao,Anping Deng,Jianguo Li
DOI: https://doi.org/10.1007/s00604-018-2790-x
IF: 5.7
2018-01-01
Microchimica Acta
Abstract:A flower-like Au/Cu alloy nanocomposite (Au/Cu NFs) was synthesized and used in an electrochemiluminescence (ECL) based method for sensitive determination of the dye Sudan I. The Au-g-C3N4 nanosheets as an ECL emitter were prepared by electrostatic adsorption between gold nanoparticles and g-C3N4. They form a film on a glassy carbon electrode (GCE) and then can be connected with Sudan I antigen via gold-nitrogen bond and amidation reactions. The Au/Cu NFs combined with Sudan I antibody also via the Au-N bond and was introduced into the modified GCE by specific recognition between the antibody and the antigen. The overlap between emission spectra of the Au-g-C3N4 nanosheets and absorption spectra of Au/Cu NFs enabled the appearance of ECL resonance energy transfer process. That is, when the Sudan I analyte not present, the ECL was weakened due to absorption by the gray Au/Cu NFs on applying voltages from −1.7 V to 0 V. Conversely, the Au/Cu NFs on the GCE are reduced due to the competition for the antibody between the analyte and the antigen. A strong green ECL emission was obtained. The ECL response is linear in the 0.5 pg mL−1 to 100 ng mL−1 Sudan I concentration range, and the detection limit is 0.17 pg mL−1.
What problem does this paper attempt to address?