A spiral-based volumetric acquisition for MR temperature imaging.

Samuel W Fielden,Xue Feng,Li Zhao,G Wilson Miller,Matthew Geeslin,Robert F Dallapiazza,W Jeffrey Elias,Max Wintermark,Kim Butts Pauly,Craig H Meyer
DOI: https://doi.org/10.1002/mrm.26981
2018-01-01
Magnetic Resonance in Medicine
Abstract:PurposeTo develop a rapid pulse sequence for volumetric MR thermometry. MethodsSimulations were carried out to assess temperature deviation, focal spot distortion/blurring, and focal spot shift across a range of readout durations and maximum temperatures for Cartesian, spiral-out, and retraced spiral-in/out (RIO) trajectories. The RIO trajectory was applied for stack-of-spirals 3D imaging on a real-time imaging platform and preliminary evaluation was carried out compared to a standard 2D sequence in vivo using a swine brain model, comparing maximum and mean temperatures measured between the two methods, as well as the temporal standard deviation measured by the two methods. ResultsIn simulations, low-bandwidth Cartesian trajectories showed substantial shift of the focal spot, whereas both spiral trajectories showed no shift while maintaining focal spot geometry. In vivo, the 3D sequence achieved real-time 4D monitoring of thermometry, with an update time of 2.9-3.3 s. ConclusionSpiral imaging, and RIO imaging in particular, is an effective way to speed up volumetric MR thermometry. Magn Reson Med 79:3122-3127, 2018. (c) 2017 International Society for Magnetic Resonance in Medicine.
What problem does this paper attempt to address?