Improved Spectrometer-Microscope for Quantitative Fluorescence Resonance Energy Transfer Measurement Based on Simultaneous Spectral Unmixing of Excitation and Emission Spectra

Fangrui Lin,Mengyan Du,Fangfang Yang,Lichun Wei,Tongsheng Chen
DOI: https://doi.org/10.1117/1.jbo.23.1.016006
IF: 3.758
2018-01-01
Journal of Biomedical Optics
Abstract:Based on our recently developed quantitative fluorescence resonance energy transfer (FRET) measurement method using simultaneous spectral unmixing of excitation and emission spectra (ExEm-spFRET), we here set up an improved spectrometer-microscope (SM) for implementing modified ExEm-spFRET (mExEm-spFRET), in which a system correction factor (fsc) is introduced. Our SM system is very stable for at least six months. Implementation of mExEm-spFRET with four or two excitation wavelengths on SM for single living cells expressing different FRET constructs obtained consistent FRET efficiency (E) and acceptor-donor concentration ratio (Rc) values. We also performed mExEm-spFRET measurement for single living cells coexpressing cyan fluorescent protein (CFP)-Bax and yellow fluorescent protein (YFP)-Bax and found that the E values between CFP-Bax and YFP-Bax were very low (2.2%) and independent of Rc for control cells, indicating that Bax did not exist as homooligomer in healthy cells, but positively proportional to Rc in the case of Rc<1 and kept constant value (25%) when Rc>1 for staurosporine (STS)-treated cells, demonstrating that all Bax formed homooligomer after STS treatment for 6 h.
What problem does this paper attempt to address?