Detection of Gaseous Elemental Mercury Using a Frequency-Doubled Green Diode Laser.

Xiutao Lou,Tie Zhang,Hongze Lin,Shiyi Gao,Lianjie Xu,Junnan Wang,Li Wan,Sailing He
DOI: https://doi.org/10.1364/oe.24.027509
IF: 3.8
2016-01-01
Optics Express
Abstract:We demonstrate a second-harmonic-generation (SHG) based method for the detection of gaseous elemental mercury by using a newly available green diode laser. Multimode ultraviolet radiation at 253.7 nm is generated through a process of SHG. Correlation spectroscopy is introduced into the scheme to guarantee the measurement accuracy. The limit of detection achieved is 0.6 μg/m3 (0.07 ppb) for 1-m pathlength and 10-s integration time. The measurement accuracy is estimated to be 1.2%. The linear response range is estimated to be 0~60 μg/m2 (6.7 ppb·m), within which the linearity error is less than 1%. Real-time monitoring of mercury volatilization is demonstrated with a time resolution of 1 s. The results of performance characterization show that the proposed method has great potentials for mercury sensing in environmental and industrial fields.
What problem does this paper attempt to address?