Simultaneously Extracting Multiple Parameters Via Multi-Distance and Multi-Exposure Diffuse Speckle Contrast Analysis.

Jialin Liu,Hongchao Zhang,Jian Lu,Xiaowu Ni,Zhonghua Shen
DOI: https://doi.org/10.1364/boe.8.004537
IF: 3.4
2017-01-01
Biomedical Optics Express
Abstract:Recent advancements in diffuse speckle contrast analysis (DSCA) have opened the path for noninvasive acquisition of deep tissue microvasculature blood flow. In fact, in addition to blood flow index αDB , the variations of tissue optical absorption μa , reduced scattering coefficients [Formula: see text], as well as coherence factor β can modulate temporal fluctuations of speckle patterns. In this study, we use multi-distance and multi-exposure DSCA (MDME-DSCA) to simultaneously extract multiple parameters such as μa , [Formula: see text], αDB , and β. The validity of MDME-DSCA has been validated by the simulated data and phantoms experiments. Moreover, as a comparison, the results also show that it is impractical to simultaneously obtain multiple parameters by multi-exposure DSCA (ME-DSCA).