Assessment of accuracy and precision of 3D reconstruction of unicompartmental knee arthroplasty in upright position using biplanar radiography.

Tsung-Yuan Tsai,Dimitris Dimitriou,Ali Hosseini,Ming Han Lincoln Liow,Martin Torriani,Guoan Li,Young-Min Kwon
DOI: https://doi.org/10.1016/j.medengphy.2016.04.002
IF: 2.356
2016-01-01
Medical Engineering & Physics
Abstract:•The 3D knee alignment was assessed using biplanar radiography (EOS).•Unicompartmental knee arthroplasty position was assessed using EOS imaging.•The precision of the 3D component position was less than 0.29mm and 0.66°.•The precision of the 3D knee alignment was less than 1.70°.•The EOS-based technique precisely determined component position and knee alignment.