A Photoacoustic Imaging System with Variable Gain at Different Depths

Tian Guan,Yao Li,Muqun Yang,Yong Jiang,Yonghong He
DOI: https://doi.org/10.1142/s1793545818500220
IF: 2.396
2018-01-01
Journal of Innovative Optical Health Sciences
Abstract:We established a photoacoustic imaging (PAI) system that can provide variable gain at different depths. The PAI system consists of a pulsed laser with an optical parametric oscillator working at a 728[Formula: see text]nm wavelength and an imaging-acquisition-and-processing unit with an ultrasound transducer. A voltage-controlled attenuator was used to realize variable gain at different depths when acquiring PAI signals. The proof-of-concept imaging results for variable gain at different depths were achieved using specific phantoms. Both resolution and optical contrast obtained through the results of variable gain for a targeted depth range are better than those of constant gain for all depths. To further testify the function, we imaged the sagittal section of the body of in vivo nude mice. In addition, we imaged an absorption sample embedded in a chicken breast tissue, reaching a maximum imaging depth of [Formula: see text]4.6[Formula: see text]cm. The results obtained using the proposed method showed better resolution and contrast than when using 50[Formula: see text]dB gain for all depths. The depth range resolution was [Formula: see text]1[Formula: see text]mm, and the maximum imaging depth of our system reached [Formula: see text]4.6[Formula: see text]cm. Furthermore, blood vessels can be revealed and targeted depth range can be selected in nude mice imaging.
What problem does this paper attempt to address?