Breath-by-breath measurement of exhaled ammonia by acetone-modifier positive photoionization ion mobility spectrometry via online dilution and purging sampling

Lu Wang,Dandan Jiang,Lei Hua,Chuang Chen,Dongming Li,Weiguo Wang,Yiqian Xu,Qimu Yang,Haiyang Li,Song Leng
DOI: https://doi.org/10.1016/j.jpha.2023.02.008
IF: 14.026
2023-04-01
Journal of Pharmaceutical Analysis
Abstract:Exhaled ammonia (NH<sub>3</sub>) is an essential noninvasive biomarker for disease diagnosis. In this study, an acetone-modifier positive photoionization ion mobility spectrometry (AM-PIMS) method was developed for accurate qualitative and quantitative analysis of exhaled NH<sub>3</sub> with high selectivity and sensitivity. Acetone was introduced into the drift tube along with the drift gas as a modifier, and the characteristic NH<sub>3</sub> product ion peak of (C<sub>3</sub>H<sub>6</sub>O)<sub>4</sub>NH<sub>4</sub><sup>+</sup> (<i>K</i><sub>0</sub> = 1.45 cm<sup>2</sup>/V·s) was obtained through the ion-molecule reaction with acetone reactant ions (C<sub>3</sub>H<sub>6</sub>O)<sub>2</sub>H<sup>+</sup> (<i>K</i><sub>0</sub> = 1.87 cm<sup>2</sup>/V·s), which significantly increased the peak-to-peak resolution and improved the accuracy of exhaled NH<sub>3</sub> qualitative identification. Moreover, the interference of high humidity and the memory effect of NH<sub>3</sub> molecules were significantly reduced via online dilution and purging sampling, thus realizing breath-by-breath measurement. As a result, a wide quantitative range of 5.87-140.92 μmol/L with a response time of 40 ms was achieved, and the exhaled NH<sub>3</sub> profile could be synchronized with the concentration curve of exhaled CO<sub>2</sub>. Finally, the analytical capacity of AM-PIMS was demonstrated by measuring the exhaled NH<sub>3</sub> of healthy subjects, demonstrating its great potential for clinical disease diagnosis.
pharmacology & pharmacy
What problem does this paper attempt to address?