Single Shot, Three-Dimensional Fluorescence Microscopy with a Spatially Rotating Point Spread Function.

Zhaojun Wang,Yanan Cai,Yansheng Liang,Xing Zhou,Shaohui Yan,Dan,Piero R. Bianco,Ming Lei,Baoli Yao
DOI: https://doi.org/10.1364/boe.8.005493
IF: 3.4
2017-01-01
Biomedical Optics Express
Abstract:A wide-field fluorescence microscope with a double-helix point spread function (PSF) is constructed to obtain the specimen's three-dimensional distribution with a single snapshot. Spiral-phase-based computer-generated holograms (CGHs) are adopted to make the depth-of-field of the microscope adjustable. The impact of system aberrations on the double-helix PSF at high numerical aperture is analyzed to reveal the necessity of the aberration correction. A modified cepstrum-based reconstruction scheme is promoted in accordance with properties of the new double-helix PSF. The extended depth-of-field images and the corresponding depth maps for both a simulated sample and a tilted section slice of bovine pulmonary artery endothelial (BPAE) cells are recovered, respectively, verifying that the depth-of-field is properly extended and the depth of the specimen can be estimated at a precision of 23.4nm. This three-dimensional fluorescence microscope with a framerate-rank time resolution is suitable for studying the fast developing process of thin and sparsely distributed micron-scale cells in extended depth-of-field.
What problem does this paper attempt to address?