A New Scanning Mode to Improve Scanning Ion Conductance Microscopy Imaging Rate with Pipette Predicted Movement
Jian Zhuang,Yangbohan Jiao,Vincent Mugabo
DOI: https://doi.org/10.1016/j.micron.2017.07.007
IF: 2.381
2017-01-01
Micron
Abstract:Scanning ion conductance microscopy (SICM) is a non-contact surface topography measurement technique that has been increasingly used for soft surfaces such as living biological samples. An approach-retract scanning (ARS) mode is widely used to avoid collision between the SICM probe (i.e., pipette) and an abrupt increase in sample profile. However, the redundant pipette trajectory in the ARS mode lengthens the scan time, thus reducing SICM efficiency and time resolution. To avoid this problem, a new scanning mode is discussed that adds horizontal movement at each measurement point to predict the upcoming sample topography via variation in ion current. The pipette then retracts in response to raised topography, while it raster scans flat or downhill topography. The feasibility was verified by finite element analysis and experimental tests on three kinds of soft samples: polydimethylsiloxane, mice cardiac fibroblasts, and breast cancer cells. The pixel detection frequency during imaging and the mean square error of the sample topography were compared for the two modes. The new scanning mode enhances the SICM imaging rate without loss of imaging quality or scanning stability, while it increases efficiency and time resolution. It thus has an improved performance for characterizing biological samples.
What problem does this paper attempt to address?