Quantifying the Accuracy and Precision of a Novel Real-Time 6 Degree-of-freedom Kilovoltage Intrafraction Monitoring (KIM) Target Tracking System

J-H Kim,D. T. Nguyen,C-Y Huang,T. Fuangrod,V. Caillet,R. O'Brien,P. Poulsen,J. Booth,P. Keall
DOI: https://doi.org/10.1088/1361-6560/aa6ed7
IF: 3.5
2017-01-01
Physics in Medicine and Biology
Abstract:Target rotation can considerably impact the delivered radiotherapy dose depending on the tumour shape. More accurate tumour pose during radiotherapy treatment can be acquired through tracking in 6 degrees-of-freedom (6 DoF) rather than in translation only. A novel real-time 6 DoF kilovoltage intrafraction monitoring (KIM) target tracking system has recently been developed. In this study, we experimentally evaluated the accuracy and precision of the 6 DoF KIM implementation. Real-time 6 DoF KIM motion measurements were compared against the ground truth motion retrospectively derived from kV/MV triangulation for a range of lung and prostate tumour motion trajectories as well as for various static poses using a phantom. The accuracy and precision of 6 DoF KIM were calculated as the mean and standard deviation of the differences between KIM and kV/MV triangulation for each DoF, respectively. We found that KIM is able to provide 6 DoF motion with sub-degree and sub-millimetre accuracy and precision for a range of realistic tumour motion.