Sensor Configuration and Algorithms for Power-Line Interference Suppression in Low Field Nuclear Magnetic Resonance
Xiaolei Huang,Hui Dong,Quan Tao,Mengmeng Yu,Yongqiang Li,Liangliang Rong,Hans-Joachim Krause,Andreas Offenhäusser,Xiaoming Xie
DOI: https://doi.org/10.3390/s19163566
IF: 3.9
2019-08-15
Sensors
Abstract:Low field (LF) nuclear magnetic resonance (NMR) shows potential advantages to study pure heteronuclear J-coupling and observe the fine structure of matter. Power-line harmonics interferences and fixed-frequency noise peaks might introduce discrete noise peaks into the LF-NMR spectrum in an open environment or in a conductively shielded room, which might disturb J-coupling spectra of matter recorded at LF. In this paper, we describe a multi-channel sensor configuration of superconducting quantum interference devices, and measure the multiple peaks of the 2,2,2-trifluoroethanol J-coupling spectrum. For the case of low signal to noise ratio (SNR) 1 for most of the signal peaks. The DWA-GD algorithm improves the SNR further, but takes more computational time. Depending on whether the accuracy or the speed of the de-noising process is more important in LF-NMR applications, the choice of algorithm should be made.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation