Sensitive Detection of Carbon Monoxide Based on a QEPAS Sensor with a 2.3μm Fiber-Coupled Antimonide Diode Laser

Yufei Ma,Guang Yu,Jingbo Zhang,Xin Yu,Rui Sun
DOI: https://doi.org/10.1088/2040-8978/17/5/055401
IF: 2.1
2015-01-01
Journal of Optics
Abstract:In this paper, a sensitive quartz-enhanced photoacoustic spectroscopy (QEPAS)-based carbon monoxide (CO) sensor using a 2.3 mu m continuous wave (CW) distributed feedback (DFB) fibercoupled antimonide diode laser is demonstrated for the first time. Wavelength modulation spectroscopy and a second-harmonic detection technique are used to reduce the sensor background noise and simplify the data process. Water vapor is added into a 1000 ppm CO:N-2 gas mixture to improve the vibrational-translational relaxation rate of the analyzed CO for the purpose of enhancement of the QEPAS signal. After the modulation depth is optimized, a minimum detection limit of 43.3 ppm at the 4294.6 cm(-1) absorption line is achieved when the modulation depth is set to 0.32 cm(-1).
What problem does this paper attempt to address?