SU-E-T-411: Evaluation of Treatment Plan Quality Between Two Treatment Planning Systems for VMAT

tsungying yang,weiqing xu,rui ge,bo qu,shengqian xu
DOI: https://doi.org/10.1118/1.4924772
IF: 4.506
2015-01-01
Medical Physics
Abstract:Purpose: To investigate the treatment plan quality between two treatment planning systems (TPS) for volumetric modulated Arc therapy (VMAT). Methods: VMAT plans based on Varian Eclipse (Ver10.0) and Philips Pinnacle (Ver9.10) TPS were respectively generated by 10 cervical cancer patients by using 2, 3 arcs, controlling the gantry movement with the space of 2 degrees and setting the same optimization constraint as soon as possible. The delivery system with VMAT was Varian Clinac iX machine. The prescription dose was 50Gy/23f for regions at high risk (PTV1), and 45Gy/23f for regions at low risk (PTV). The difference of treatment plans was evaluated by dose distribution and delivery efficiency. Results: The VMAT 2-arcs plans based on Pinnacle in the conformal indexes of PTV(CI: 0.77±0.05 vs 0.75±0.02, p>0.05), and OARs (such as V30, V40 of rectum and bladder) showed slightly better than those of Eclipse TPS, and the homogeneity indexes (HI) of in Pinnacle were slightly worse than the Eclipse (PTV: 0.17±0.01 vs 0.14±0.01, p 0.05) were more than those of 2-arcs. Conclusion: For patients with cervical cancer, the VMAT treatment plans of two TPSs could achieve clinical acceptable dose distribution. However, we will still need more cases to further study and determine the optimization and performance characteristics of the two commercial TPSs.