A Level Set Method for Cupping Artifact Correction in Cone-Beam CT.

Shipeng Xie,Chunming Li,Haibo Li,Qi Ge
DOI: https://doi.org/10.1118/1.4926851
IF: 4.506
2015-01-01
Medical Physics
Abstract:PURPOSE:To reduce cupping artifacts and improve the contrast-to-noise ratio in cone-beam computed tomography (CBCT).METHODS:A level set method is proposed to reduce cupping artifacts in the reconstructed image of CBCT. The authors derive a local intensity clustering property of the CBCT image and define a local clustering criterion function of the image intensities in a neighborhood of each point. This criterion function defines an energy in terms of the level set functions, which represent a segmentation result and the cupping artifacts. The cupping artifacts are estimated as a result of minimizing this energy.RESULTS:The cupping artifacts in CBCT are reduced by an average of 90%. The results indicate that the level set-based algorithm is practical and effective for reducing the cupping artifacts and preserving the quality of the reconstructed image.CONCLUSIONS:The proposed method focuses on the reconstructed image without requiring any additional physical equipment, is easily implemented, and provides cupping correction through a single-scan acquisition. The experimental results demonstrate that the proposed method successfully reduces the cupping artifacts.
What problem does this paper attempt to address?