Intercomparison of air ion spectrometers: a basis for data interpretation

s gagne,k lehtipalo,h e manninen,tuomo nieminen,siegfried schobesberger,alessandro franchin,t ylijuuti,j boulon,a sonntag,s mirme,aadu mirme,urmas horrak,t petaja,e asmi,markku kulmala
DOI: https://doi.org/10.5194/amtd-4-1139-2011
2011-01-01
Atmospheric Measurement Techniques Discussions
Abstract:Abstract. We evaluated 11 air ion spectrometers from Airel Ltd. after they had spent one year in field measurements as a part of the EUCAARI project: 5 Air Ion Spectrometers (AIS), 5 Neutral cluster and Air Ion Spectrometers (NAIS) and one Airborne NAIS (ANAIS). This is the first time that an ANAIS is evaluated and compared so extensively. The ion spectrometers' mobility and concentration accuracy was evaluated. Their measurements of ambient air were compared between themselves and to reference instruments: a DMPS, a BSMA, and an Ion-DMPS. We report on the simultaneous measurement of a new particle formation (NPF) event by all 11 instruments and the 3 reference instruments. To our knowledge, it is the first time that the size distribution of ions and particles is measured by so many ion spectrometers during a NPF event. The new particle formation rates (~ 0.2 cm−3 s−1 for ions and ~ 2 cm−3 s−1 for particles) and growth rates (~ 25 nm h−1 in the 3–7 nm size range) were calculated for all the instruments. The NAISs and the ANAIS gave higher concentrations and formation rates than the AISs. The latter agreed better with reference instruments. Finally, based on the results presented here, we give guidelines for data interpretation, when data from different ion spectrometers are compared.
What problem does this paper attempt to address?