Monitoring of Permeability of Different Analytes in Human Normal and Cancerous Bladder Tissues in Vitro Using Optical Coherence Tomography

Bingsong Lei,Xiaoyuan Deng,Huajiang Wei,Guoyong Wu,Zhouyi Guo,Hongqin Yang,Yonghong He,Shusen Xie
DOI: https://doi.org/10.1070/qe2014v044n12abeh015380
2014-01-01
Quantum Electronics
Abstract:We report our preliminary results on quantification of glucose and dimethyl sulfoxide (DMSO) diffusion in normal and cancerous human bladder tissues in vitro by using a spectral domain optical coherence tomography (SD-OCT). The permeability coefficients (PCs) of a 30% aqueous solution of glucose are found to be (7.92 +/- 0.81) x 10(-6) cm s(-1) and (1.19 +/- 0.13) x 10(-5) cm s(-1) in normal and cancerous bladder tissues, respectively. The PCs of 50% DMSO are calculated to be (8.99 +/- 0.93) x 10(-6) cm s(-1) and (1.43 +/- 0.17) x 10(-5) cm s(-1) in normal and cancerous bladder tissues, respectively. The obtained results show a statistically significant difference in permeability of normal and cancerous tissue and indicate that the PC of 50% DMSO is about 1.13- and 1.21-fold higher than that of 30% glucose in normal bladder and cancerous bladder tissues, respectively. Thus, the quantitative measurements with the help of PCs from OCT images can be a potentially powerful method for bladder cancer detection.
What problem does this paper attempt to address?