Glucose biosensing based on glucose oxidase immobilization on carboxymethyl chitosan/polyaniline/multi-walled carbon nanotubes nanocomposite
Niloofar Sadat Hadian,Hassan Faridnouri,Ehsan Nazarzadeh Zare
DOI: https://doi.org/10.1016/j.diamond.2024.111423
IF: 3.806
2024-07-20
Diamond and Related Materials
Abstract:A novel electrochemical glucose biosensor was fabricated using the casting technique of multi-walled carbon nanotubes (MWCNT)/polyaniline (PANI)/carboxymethyl chitosan (CCs) as an appropriate matrix for glucose oxidase (GOx) immobilization. The differential pulse voltammetry (DPV) study results demonstrated a wide linear range of glucose concentrations from 10 nM to 10 μM, with a limit of detection (LOD) of 1.41 μM and a sensitivity of 1791 μA mM −1 cm −2 . Additionally, cyclic voltammetry (CV) data analysis yielded a heterogeneous rate constant (ks) of 0.3 s −1 and an apparent Michaelis-Menten constant ( kmapp ) of 18 nM. Over a 30-day period, the biosensor exhibited exceptional repeatability and stability, maintaining approximately 84.58 % of its initial performance for storage stability at 4 °C and 82.13 % for functional stability. Moreover, the reproducibility, interference resistance, and overall performance of the biosensor were assessed, demonstrating its capability to accurately measure glucose levels in human serum with a relative standard deviation (RSD) of <5 %.
materials science, multidisciplinary,physics, applied, condensed matter, coatings & films