Analysis of Surface Absorbed Dose in X-ray Grating Interferometry

Zhili Wang,Zhao Wu,Kun Gao,Dajiang Wang,Heng Chen,Shenghao Wang,Ziyu Wu
DOI: https://doi.org/10.1016/j.elspec.2014.04.010
IF: 1.993
2014-01-01
Journal of Electron Spectroscopy and Related Phenomena
Abstract:X-ray phase contrast imaging using grating interferometry has shown increased contrast over conventional absorption imaging, and therefore the great potential of dose reduction. The extent of the dose reduction depends on the geometry of grating interferometry, the photon energy, the properties of the sample under investigation and the utilized detector. These factors also determine the capability of grating interferometry to distinguish between different tissues with a specified statistical certainty in a single raw image. In this contribution, the required photon number for imaging and the resulting surface absorbed dose are determined in X-ray grating interferometry, using a two-component imaging object model. The presented results confirm that compared to conventional radiography, phase contrast imaging using grating interferometry indeed has the potential of dose reduction. And the extent of dose reduction is strongly dependent on the imaging conditions. Those results provide a theoretical framework for dose estimation under given imaging conditions before experimental trials, and general guidelines for optimization of grating interferometry for those dose-sensitive applications. (C) 2014 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?