A novel all-solid-state ammonium electrode with polyaniline and copolymer of aniline/2,5-dimethoxyaniline as transducers

yuanfeng huang,jun li,tianya yin,jianjun jia,qian ding,hao zheng,chentung arthur chen,ying ye
DOI: https://doi.org/10.1016/j.jelechem.2014.12.041
IF: 4.598
2015-01-01
Journal of Electroanalytical Chemistry
Abstract:•In this study we reported an all-solid-state ammonium ion selective electrode.•Polyaniline and copolymer of aniline/2,5-dimethoxyaniline were used as transducer.•The dime-size electrode is easy to be integrated with diameter of 0.6mm.•Its response time is <1 second, lifetime is >3months.•It was suitable for in-situ monitoring of NH4+ in pH range of 2.6–10.1.